AEROCHAMBER* VENT Holding Chamber

For use with an MDI for mechanically ventilated patients
Asthma Management
Bronchiectasis
COPD
Cystic Fibrosis
Inhaler

The AEROCHAMBER* VENT Holding Chamber is used to deliver medication from metered dose inhalers (MDIs) to mechanically ventilated patients. The chamber can be left in the inspiratory limb of a ventilator circuit; it can be used for medication delivery and then collapsed without interrupting continuous mechanical ventilation.

  • Clinically validated for use in ventilator circuits1
  • Maintains positive end expiratory pressure (PEEP) when installed in the ventilator circuit
  • Design is simple to connect, open and close with less disruption to the patient

1. Alexander G Duarte, et al . Bronchodilator Therapy With Metered-Dose Inhaler And Spacer Versus Nebulizer In Mechanically Ventilated Patients: Comparison Of Magnitude And Duration Of Response. Respiratory Care 2000;45(7):817-823.                                             

Currently this is not available for purchase through this website.

Why use an inhaler always together with an AEROCHAMBER* VENT spacer?

Use of a spacer simplifies coordination of MDI actuation and inhalation, reducing aerosol losses with poor hand-breath coordination.2

2. Douglas S Gardenhire, et al. A Guide To Aerosol Delivery Devices for Respiratory Therapists. 5th Edition. American Association for Respiratory Care 2023.


Features

  • Collapsible to minimize water trapping
  • AERODOCK* Canister Port and a removable canister adaptor allows for use with all types of metered dose inhalers, including those with integrated dose counters
  • Offset wye connector, for use in both standard and parallel wye circuits
  • Single-patient, multiple use (replace with the circuit)

Download instructions for use for additional information

labelled line drawing of aerochamber vent holding chamber

A Cap | B Canister Adapter | C Canister Port | D Latch Release Button | E Bellow | F Catch | G Wye Connector

Image
diagram depicting how to check if cap and canister adapter are securely in place before installing ventilator circuit

Step 1

Ensure the cap (A) and canister adapter (B) are securely in place in the canister port (C) prior to installing in the ventilator circuit.

Image
diagram depicting how to remove inspiratory limb from the wye of aerochamber vent holding chamber

Step 2

Disconnect the inspiratory limb of the circuit from the wye. Connect the wye connector (G) of the chamber to the wye. Connect the other end of the chamber to the inspiratory limb of the circuit. The canister port (C) must be on top and distal to the patient.

Image
diagram depicting how to latch chamber in collapsed position of aerochamber vent holding chamber

Step 3

Latch the chamber in the collapsed position by holding the patient end stationary while gently pushing the canister port (C) end towards it until you hear the latch release buttons (D) “click”.

Image
diagram depicting how to ensure aerochamber vent holding chamber device is fully collapsed with latches engaged

Step 4

Ensure that the device is fully collapsed and that the catches (F) are completely engaged by lightly pulling the ends away from each other. The device should remain latched, if not repeat the latching step.

Use Instructions to Administer Medication

Image
diagram depicting how to unlatch and expand aerochamber vent holding chamber

Step 1A

Unlatch and expand the chamber by: holding the wye connector (G) close to the wye.

Image
diagram depicting how to position thumb and index finger aerochamber vent holding chamber

Step 1B

Position thumb and index finger of the other hand on the latch release buttons (D) and press together until they disengage.

Image
diagram depicting how to hold aerochamber vent holding chamber

Step 1C

Hold the patient end stationary and gently pull the canister end away from the patient end until the bellow (E) is fully expanded.

Image
diagram depicting how to use aerochamber vent holding chamber for MDI canisters

Step 2A

For Conventional MDI Canisters: Remove the cap (A) from the canister port (C). Ensure the canister adapter (B) remains inside the canister port (C). Shake the MDI canister immediately before use as per the instructions supplied with the MDI. Insert the MDI canister into the canister port (C), being careful not to discharge the canister at this time.

Image

Step 2B

For MDI Canisters With Integrated Dose Counters (GlaxoSmithKline† type): Remove both the cap (A) and the canister adapter (B) from the canister port (C). Shake the MDI canister immediately before use as per the instructions supplied with the MDI. Insert the MDI canister with integrated dose counter into the canister port (C), being careful not to discharge the canister at this time. The integrated counter will be functional upon actuation.

Image
diagram depicting how to hold aerochamber vent holding chamber to administer medication

Step 3

Hold the chamber in your hand maintaining a straight horizontal position. Hold the canister in a vertical position. Prior to the beginning of inspiration, actuate the canister.

Image
diagram depicting how to remove mdi canister from aerochamber vent holding chamber

Step 4

Following treatment, remove the MDI canister from the chamber. Replace the canister adapter (B) (if removed) and cap (A).

Image
diagram depicting how to latch chamber in collapsed position after using aerochamber vent holding chamber

Step 5

Latch the chamber in the collapsed position by holding the patient end stationary while gently pushing the canister port (C) end towards it until you hear the latch release buttons (D) “click”.

Image
diagram depicting how to ensure aerochamber vent holding chamber device is fully collapsed

Step 6

Ensure that the device is fully collapsed and that the catches (F) are completely engaged by lightly pulling the ends away from each other. The device should remain latched, if not repeat the latching step.

Cleaning

The device should not be cleaned. It can remain in the ventilator circuit and replaced along with the ventilator circuit.

Refer to the instructions for use for additional information.

Download

 

Product Information


AEROCHAMBER* VENT Holding Chamber Instructions for Use

Instructions for use.

Download


AEROCHAMBER* VENT Holding Chamber Product Information Sheet

Refer to the instructions for use for additional information.

Download

Scientific & Performance Evidence


Does Spacer/Adapter Device Choice Affect Delivery of a Pressurized Metered Dose Inhaler (pMDI) through a Humidified Circuit to a Simulated Patient on Mechanical Ventilation

AEROCHAMBER* VENT Holding Chamber delivered more drug during simulated adult mechanical ventilation compared with 3 other in-line devices.

Download


How Do Different Methods Of Humidification Affect Drug Delivery To A Simulated Patient Receiving Mechanical Ventilation

AEROCHAMBER* VENT Holding Chamber delivered the largest dose of salbutamol using either a traditional heated humidifier or a heat and moisture exchange (HME) filter.

Download


In Vitro Performance Of a New Collapsible Holding Chamber For The Delivery Of Aerosolized Medication By Pressurized Metered-Dose Inhaler

AEROCHAMBER* VENT Holding Chamber delivers significantly more medication to the distal end of the endotracheal tube compared with other adaptors.

Download


Bronchodilator Therapy With Metered-Dose Inhaler And Spacer Versus Nebulizer In Mechanically Ventilated Patients: Comparison Of Magnitude And Duration Of Response

AEROCHAMBER* VENT Holding Chamber is clinically validated for use in mechanical ventilator circuits.

Download


Specialty Chambers Study Summary

A study compilation covering the full range of specialty application chambers that have been designed to deliver metered dose inhaler (MDI) aerosol medication within the hospital environment.

Download