The AEROTRACH PLUS* Anti-Static Valved Holding Chamber (VHC) is specifically designed for the delivery of metered dose inhaler (MDI) medications for tracheostomy patients.
The AEROTRACH PLUS* Anti-Static Valved Holding Chamber (VHC) is quick and easy to comfortably connect to standard 15mm tracheostomy tubes.
Use of a chamber improves delivery of medication1
Ready to use directly out of the package – time saving with no pretreatment needed
Anti-static improves the reliability of medication delivery2
Remove the cap from the inhaler. Shake the inhaler immediately before use as per the instructions supplied with it.
Image
Step 2
Insert the inhaler into the backpiece of the chamber.
Image
Step 3
Connect the trach adapter of the chamber to the connection on the tracheostomy tube. Note: This device is designed as a slip fit onto the tracheostomy tube connection. It must be held in place during administration of medication.
Image
Step 4
Press the inhaler once at the beginning of a slow inhalation. Hold in place for 5-6 breaths. Administer one puff at a time. Follow the instructions supplied with the inhaler on how long to wait before repeating.
Cleaning
This anti-static chamber can be used right out-of-package and then cleaned daily. Follow the steps outlined below or download our full instructions.
Image
Step 1
Remove back piece only. (Do not remove the trach adapter assembly.)
Image
Step 2
Soak the parts for 15 minutes in a mild solution of liquid dish detergent and lukewarm clean water. Agitate gently. Rinse parts in clean water.
Image
Step 3
Shake out excess water and allow to air dry in a vertical position. Ensure parts are dry before reassembly.
Image
Step 4
Replace backpiece when unit is completely dry. Store in a clean, dry place when not in use.
Disinfection
Prior to disinfection, follow cleaning instructions. Do not dry or reassemble the device.
Step 1
Soak the two parts in a 1:10 bleach to water solution for 30 minutes. Rinse parts in clean water. Note: You may substitute an enzyme solvent or VitalOxide† for the bleach solution. Follow the disinfectant instructions for use.
Step 2
Shake out all excess water and allow to air dry in an upright position. Ensure the parts are dry before reassembly.
Step 3
Replace backpiece when unit is completely dry.
Refer to the instructions for use for additional information.
AEROTRACH PLUS* Anti-Static VHC with tracheostomy adaptor provides a reliable means of delivering soft mist inhaler (SMI) aerosols to patients with a tracheostomy tube or tracheal stoma.
Performance Of A Valved Holding Chamber (VHC) With Tracheostomy Adapter: Aerosol Delivery From Soft Mist Inhalers (SMIs)
AEROTRACH PLUS* Anti-Static VHC with tracheostomy adaptor provides a reliable means of delivering soft mist inhaler (SMI) aerosols to patients with a tracheostomy tube or tracheal stoma.
A study compilation covering the full range of specialty application chambers that have been designed to deliver metered dose inhaler (MDI) aerosol medication within the hospital environment.