AEROCHAMBER* MV Holding Chamber

For use with an MDI for mechanically ventilated patients
Asthma Management
Bronchiectasis
COPD
Cystic Fibrosis
Inhaler

The AEROCHAMBER* MV Holding Chamber can be used for adults or in paediatric applications in a ventilator circuit or with a resuscitation bag. This chamber is removed from the circuit following treatment.

  • Clinically validated for use in ventilator circuits1
  • Widely available chamber has been on the market for over 20 years

1. E Rieger-Fackeldey, et al. Effects Of Inhaled Formoterol Compared With Salbutamol In Ventilated Preterm Infants. Pulmonary Pharmacology & Therapeutics 2004;17(5):293-300.

Currently this is not available for purchase through this website.

Why use an inhaler always together with an AEROCHAMBER* MV spacer?

Use of a spacer simplifies coordination of MDI actuation and inhalation, reducing aerosol losses with poor hand-breath coordination.2

2. Douglas S Gardenhire, et al. A Guide To Aerosol Delivery Devices for Respiratory Therapists, 5th Edition. American Association for Respiratory Care 2023.


Features

  • Can be used with most MDI canisters (excluding those with integrated dose counters)
  • Clear chamber allows for visualization of aerosol
  • Single-patient use, multiple use (replace with the circuit)
labelled diagram of aerochamber mv device

A Adapter | B 15 mm Outer Diameter | C 22 mm Inner Diameter | D Canister Port

Image
Aerochamber MV-Product-Illustration-Ventilator-Circuit-How-To-Use-1

Step 1

Attach the adapter(A) to the 15 MM outer diameter (B) end of the chamber.

Image
Aerochamber MV-Product-Illustration-Ventilator-Circuit-How-To-Use-2

Step 2

Disconnect the inspiratory limb of the circuit from the wye. Connect the 22 MM inner diameter (C) end to the wye. Connect the adapter (A) to the inspiratory limb of the circuit. The canister port (D) must be on top and distal to the patient.

Image
Aerochamber MV-Product-Illustration-Ventilator-Circuit-How-To-Use-3

Step 3

Shake the MDI canister immediately before use as per the instructions supplied with the MDI. Insert the MDI canister into the canister port (D), being careful not to discharge the canister at this time.

Image
erochamber MV-Product-Illustration-Ventilator-Circuit-How-To-Use-4.

Step 4

Hold the canister in a vertical position. Prior to the beginning of inspiration, actuate the canister.

Image
Aerochamber MV-Product-Illustration-Ventilator-Circuit-How-To-Use-5

Step 5

Following treatment, remove the MDI canister from the chamber. Remove the chamber from the circuit. Reconnect the inspiratory limb of the circuit to the wye. Do not leave the chamber in the circuit after administering medication, the chamber will collect condensate water. Store in a clean, dry place when not in use.


Cleaning

The device should not be cleaned. It can be replaced with the ventilator circuit. Refer to the instructions for use for additional information.

labelled diagram of aerochamber mv device with resuscitation bag

A Adapter | B 15 mm Outer Diameter | C 22 mm Inner Diameter | D Canister Port

Image
Aerochamber MV-Product-Illustration-Resucitation-How-To-Use-1

Step 1

Attach the adapter (A) to the 22 MM inner diameter (C) end of the chamber.

Image
Attach the ADAPTER (A) to the 22 MM INNER DIAMETER (C) end of the chamber

Step 2

Attach the resuscitation bag to the 15 MM outer diameter (B) end of the chamber.

Image
Aerochamber MV-Product-Illustration-Resucitation-How-To-Use-3

Step 3

Disconnect the wye from the endotracheal tube. Connect the adapter (A) to the endotracheal tube. The canister port (D) must be on top.

Image
Aerochamber MV-Product-Illustration-Resucitation-How-To-Use-4

Step 4

Shake the MDI canister immediately before use as per the instructions supplied with the MDI. Insert the MDI canister into the canister port (D), being careful not to discharge the canister at this time.

Image
Aerochamber MV-Product-Illustration-Resucitation-How-To-Use-5

Step 5

Hold the canister in a vertical position. Actuate the canister and immediately deliver a manual breath.

Image
Aerochamber MV-Product-Illustration-Resucitation-How-To-Use-6

Step 6

Following treatment, remove the MDI canister from the chamber. Disconnect the chamber from the endotracheal tube. Reconnect the wye to the endotracheal tube. Store in a clean, dry place when not in use.


Cleaning

The device should not be cleaned. It can be replaced with the ventilator circuit. Refer to the instructions for use for additional information.

Scientific & Performance Evidence

Trudell Medical International has an enviable history of strong leadership in creating innovative medical devices that enhance the quality of life for people of all ages - all backed by science and clinical evidence.


Effects Of Inhaled Formoterol Compared With Salbutamol In Ventilated Preterm Infants

AEROCHAMBER* MV Holding Chamber is clinically validated for use in mechanical ventilator circuits.

Download


Study Summary Specialty Chambers

A study compilation covering the full range of specialty application chambers that have been designed to deliver metered dose inhaler (MDI) aerosol medication within the hospital environment.

Download

Product Information


AEROCHAMBER* MV Holding Chamber Product Information Sheet

Refer to the instructions for use for additional information.

Download


AEROCHAMBER* MV Holding Chamber Instructions for Use

Instructions for use.

Download

Recommended Solutions

AeroChamber Mini angled horizontal, viewed from the left

AEROCHAMBER* mini Anti-Static Holding Chamber

For use with an MDI for mechanically ventilated patients

Small volume chamber used in a mechanical ventilator circuit or with a resuscitation bag.

AeroChamber VENT in the closed position, viewed from the top right

AEROCHAMBER* VENT Holding Chamber

For use with an MDI for mechanically ventilated patients

Can be left in the inspiratory limb of a 22mm circuit.