For use with an MDI for mechanically ventilated patients
Asthma Management
Bronchiectasis
COPD
Cystic Fibrosis
Inhaler
The AEROCHAMBER* MV Holding Chamber can be used for adults or in paediatric applications in a ventilator circuit or with a resuscitation bag. This chamber is removed from the circuit following treatment.
Clinically validated for use in ventilator circuits1
Widely available chamber has been on the market for over 20 years
A Adapter | B 15 mm Outer Diameter | C 22 mm Inner Diameter | D Canister Port
Image
Step 1
Attach the adapter(A) to the 15 MM outer diameter (B) end of the chamber.
Image
Step 2
Disconnect the inspiratory limb of the circuit from the wye. Connect the 22 MM inner diameter (C) end to the wye. Connect the adapter (A) to the inspiratory limb of the circuit. The canister port (D) must be on top and distal to the patient.
Image
Step 3
Shake the MDI canister immediately before use as per the instructions supplied with the MDI. Insert the MDI canister into the canister port (D), being careful not to discharge the canister at this time.
Image
Step 4
Hold the canister in a vertical position. Prior to the beginning of inspiration, actuate the canister.
Image
Step 5
Following treatment, remove the MDI canister from the chamber. Remove the chamber from the circuit. Reconnect the inspiratory limb of the circuit to the wye. Do not leave the chamber in the circuit after administering medication, the chamber will collect condensate water. Store in a clean, dry place when not in use.
A Adapter | B 15 mm Outer Diameter | C 22 mm Inner Diameter | D Canister Port
Image
Step 1
Attach the adapter (A) to the 22 MM inner diameter (C) end of the chamber.
Image
Step 2
Attach the resuscitation bag to the 15 MM outer diameter (B) end of the chamber.
Image
Step 3
Disconnect the wye from the endotracheal tube. Connect the adapter (A) to the endotracheal tube. The canister port (D) must be on top.
Image
Step 4
Shake the MDI canister immediately before use as per the instructions supplied with the MDI. Insert the MDI canister into the canister port (D), being careful not to discharge the canister at this time.
Image
Step 5
Hold the canister in a vertical position. Actuate the canister and immediately deliver a manual breath.
Image
Step 6
Following treatment, remove the MDI canister from the chamber. Disconnect the chamber from the endotracheal tube. Reconnect the wye to the endotracheal tube. Store in a clean, dry place when not in use.
Trudell Medical International has an enviable history of strong leadership in creating innovative medical devices that enhance the quality of life for people of all ages - all backed by science and clinical evidence.
Effects Of Inhaled Formoterol Compared With Salbutamol In Ventilated Preterm Infants
AEROCHAMBER* MV Holding Chamber is clinically validated for use in mechanical ventilator circuits.
A study compilation covering the full range of specialty application chambers that have been designed to deliver metered dose inhaler (MDI) aerosol medication within the hospital environment.